- Easy to Use; Take It Into the Shop, Plant or Field
- Simulates 1 Type, Pt100, 0.00385
- Calibrate Directly in C/F
- Compatible With All Process Instruments
- Nist Certificate Included
- Accurate to +/-0.25C (+/-0.5F) for Pt 100 Ohms RTD's w/0.1C Resolution
- Guaranteed Compatible w/Devices Using Pulsed Excitation Currents Including PLCs, DCS, Multi Channel Recorders
- EZ-DIAL Knob With 2 Dialing Speeds Makes Adjusting Output Easy
- EZ-Check Switch Stores Settings for 0% & 100% Span Adjustments
- Battery Life 45 Hrs Under Typical Continuous Usage
The 510B-PT-100-1 simulates RTD resistances into all types of process instruments including transmitters, recorders, controllers, alarms, data aquisition, and computer systems. No longer will you need to lug a decade box around or make errors reading RTD tables incorrectly. The 510B-PT-100-1 is a superior replacement for decade boxes by allowing you to dial directly in temperature. It also has better accuracy than many higher priced calibrators.
With the 510B-PT-100-1 you can continuously select output temperatures. And you can instantly select 3 outputs ranges by flipping the EZ-Check switch to calibrate zero, span and mid-point. Built-in LCD display with backlight helps for easy viewing in dark locations.
To Order the model below scroll to the top of the page then click on ADD TO CART button |
PART NUMBER | DESCRIPTION |
CL-PI-510B-PT-100-1 | RTD Simulator, 1 Type (Pt100, 0.00385) |
Comes with NIST Cert, 4-AA alkaline batteries, 2 red & 2 black leads w/retracable shield banana plugs & spade lugs.
SPECIFICATIONS (unless otherwise indicated all specs are rated from a nominal 23C, 70% RH for 1 year from calibration)
ACCURACY: +/-(0.015% of setting in ohms + 0.05 ohms)
TEMPERATURE DRIFT: +/- 0.05 ohms/C
OPERATING TEMP RANGE: -25 to 60C (-10 to 140F)
RH RANGE: 10% <RH <90% (0 to 35C) non-condensing
DIMENSIONS: 4.96 x 2.73 x 1.79 in (126 x 69 x 45 mm)
WEIGHT: 8.4 oz (0.24 kg (including 4 AA batteries)
BATTERY LIFE: 45 hrs with low battery warning of 1 hr left
PROTECTION AGAINST MISCONNECTION: Over-voltage protection to 60 V dc (rated for 30 seconds)
DISPLAY: High contrast LCD with LED backlight for use in dim lit areas
SIMULATE
Accuracy from 1 to 10.2 mA external excitation current: +/-(0.015% of setting + 0.005 ohms)
Accuracy below 1 mA of external excitation: +/-(0.015% of setting +0.01mV/mA excitation current + 0.05 ohms)
Resistance ranges, 0.00 to 410.00, 410.1 to 4001.0 ohms
Allowable excitation current range, <410 ohms: 10.2 mA max; steady or pulsed/intermittent
Pulsed excitation current compatibility DC to 0.01 seconds pulse width