- Heated Chilled Mirror System for Elevated Dew Points Above Ambient
- Temperature Controlled 2 Stage Chilled Mirror Hygrometer
- Measures Elevated Dew Points Up to 95°C in Non-Condensing Environments
- Dew Point Measurement Range from -30°C to 95°C
- Accurate Measurement of Dew Point/Frost Point in a Gas Sample
- Very Responsive & Repeatable to +/-0.05°C
- Bi-directional RS232 Communication
- NIST Traceable Certificate Included
The CM-RO-DEWTRAN-HB is a temperature controlled 2 stage chilled mirror hygrometer suited to measure elevated dew points up to 95°C in a non-condensing environment. Dew point measurement range is from -30°C to 95°C. The chilled mirror uses a fundamental principle to accurately measure the dew point/frost point in a gas sample. It is typically used as a reference standard for dew point over the measurement range. It is very responsive and repeatable to +/-0.05°C and easy to maintain.
The CM-RO-DEWTRAN-HB has a built in automatic reference cycle (ARC) method that balances the optical signal for temperature change and mirror contaminants. The unit includes a heated stainless steel sintered filter and is rated to filter particles down to 7 microns. The filter element is easy to remove and clean.
The temperature inside the sensor housing is controlled via the RS232 communications. Unit comes with a 10 ft heated hose and NIST Certificate.
To Order the models below scroll to the top of the page and choose the desired model from the "Power" drop down menu, then ADD TO CART |
PART NUMBERS | DESCRIPTION |
CM-RO-DEWTRAN-HB-115V | HEATED CHILLED MIRROR SYSTEM 115V |
CM-RO-DEWTRAN-HB-230V | HEATED CHILLED MIRROR SYSTEM 230V |
Units come with 10 ft heated hose, NIST Certificate, Manual
SPECIFICATIONS
TECHNOLOGY: chilled mirror, 2 stage TEC
CALIBRATION: NIST traceable calibration
ACCURACY: +/- 0.2C (+/-0.36F)
REPEATABILITY: +/- 0.05C (+/-0.09F)
INTERNAL HEATED TEMP RANGE: ambient to 99C (77 to 210F) settable
OPERATING TEMPERATURE: 0C to 60C
PRESSURE: max 4.5 bar
SIGNAL OUTPUT: 4-20 mA, 0-5Vdc or 0-10Vdc
COMMUNICATION: RS232 bi-directional
ALARM: 2 NC/SPDT 1 form C-0.5a 120/240 Vac
POWER: 90-265 Vac
SAMPLE FLOW: 0.5 to 3 SCFH (0.25 to 1.42 I/M)
DISPLAY: 2-line LCD (backlit)
SENSOR ELEMENT: 1/3 Class A Din, 100 ohms 0C (32F)
HOUSING: IP65 enclosure
SIZE: 10.5 x 8.25 x 10 inches
FILTER: SS sinter filter, 7 microns
CONNECTION: 1/4 inch compression