FTK Simple Fixed Mount Blackbody Temperature Reference/Checker for Inf

Free Shipping! - USA Orders Over $75

Customer Reviews
FTK Simple Fixed Mount Blackbody Temperature Reference/Checker for Infrared Temperature Sensors

FTK Simple Fixed Mount Blackbody Temperature Reference/Checker for Infrared Temperature Sensors

Regular price $1,982.00 Sale SKU:IR-CA-FTK35
Ships in 4-5 weeks

  • Quick and Accurate Way to Check the Calibration of IR Pyrometers
  • Fixed Temperature from 35°C to 150°C - Depending on Model
  • Outstanding Temperature Stability, Less than ±0.2°C Deviation
  • LED Temperature Indication

The FTK provides a quick and accurate way to check the calibration of infrared temperature sensors. These rugged and portable units are designed to provide fast calibration checks anywhere they are needed, from the factory to the workshop or laboratory. Eighteen models are available; offering target temperatures from 35°C or 150°C, all providing outstanding stability with less than ±0.2°C deviation.

The FTK takes between 5 and 15 minutes to heat and stabilize at the desired temperature (depending on the model), and uses a clear LED to show when it has reached that temperature: Green when the FTK is warming up, Orange when the FTK is ready for operation and Red when the FTK is above the calibration temperature.

The FTK can be used with any infrared temperature sensor that is able to measure between 35°C and 150°C and can focus on a target area less than 50.8 mm diameter.

 To Order the model below scroll to the top of the page then  click on ADD TO CART button
Model Target Temperature
IR-CA- 35 35 °C
IR-CA- 45 45 °C
IR-CA- 50 50 °C
IR-CA- 55 55 °C
IR-CA- 60 60 °C
IR-CA- 65 65 °C
IR-CA- 70 70 °C
IR-CA- 75 75 °C
IR-CA- 80 80 °C
IR-CA- 85 85 °C
IR-CA- 90 90 °C
IR-CA- 95 95 °C
IR-CA- 100 100 °C
IR-CA- 110 110 °C
IR-CA- 120 120 °C
IR-CA- 130 130 °C
IR-CA- 140 140 °C
IR-CA- 150 150 °C


 Downloads
FTK Data Sheet


General Specifications
Target Temperature: from 35 °C to 150 °C depending on model (see Ordering)
Emissivity: 0.98 ± 0.004 (for wavelength of 2 to 5.4 μm and 8 to 14 μm)
Aperture Diameter: 50.8 mm
Warm-up Time: 5 minutes (FTK35) to 15 minutes (FTK150) max.
Temperature Uncertainty: 0.4 °C for ambient temp. 10 to 30 °C (FTK35 - 120)
0.6 °C for ambient temp. 0 to 10 °C (FTK35 - 120)
0.5 °C for ambient temp. 10 to 40 °C (FTK130 - 150)
0.7 °C for ambient temp. 0 to 10 °C (FTK130 - 150)
Repeatability: 0.2°C
Stability: 0.1°C
Temperature Uniformity: 0.2°C (central area ø 45 mm)
Storage Temperature: 0 to 70°C
Relative Humidity: 10 to 85 %, non condensing
Status: LED
Green: warm-up
Orange: ready for operation
Red: above calibration temperature
Power Supply: 24 V DC, 1 A max.
Protection Class: IP50 (EN 60529)
Weight: 0.9 kg
Dimensions: 64.5 x 81.0 x 133.5mm (diameter x D x H)
CE marking: according to EU regulations
Power Adapter:  100 to 240 V AC, 50 Hz