ThermoVault - Oven Temperature Profiling System

Free Shipping! - USA Orders Over $75

Customer Reviews
ThermoVault - Oven Temperature Profiling System
ThermoVault - Oven Temperature Profiling System

ThermoVault - Oven Temperature Profiling System

Regular price $1,049.00 Sale SKU:ThermoVault
Ships in 2-3 weeks

  • Thermal insulation for high-temperature oven environments
  • Stand-alone, battery-powered thermocouple data logger
  • Records up to 1,047,552 readings with high precision
  • Programmable start/stop functions with user-friendly software
  • Digital calibration and optional password protection
  • Real-time clock for precise time and date stamping
  • Operates effectively up to 350 °C (662 °F) for limited durations
  • Compatible with MadgeTech 4 Software for advanced data analysis

The ThermoVault Oven Temperature Profiling System is designed for extreme temperature monitoring and data logging within industrial oven environments. This robust system consists of a TC101A-ST data logger encased in a thermal barrier, capable of enduring high temperatures while maintaining measurement precision.

Equipped with advanced features such as multiple start/stop functions, the ThermoVault ensures flexibility and ease of use. Its storage capacity allows for over one million readings per channel, providing comprehensive data coverage for extended periods. The real-time clock ensures all logs are accurately time-stamped, securing data integrity even when the battery is low.

Durability and Performance
The thermal enclosure enables the ThermoVault to withstand environments up to 350 °C (662 °F) for up to 30 minutes, making it ideal for oven profiling, powder coating, and food processing applications. The non-volatile solid-state memory ensures data is preserved, even if the battery depletes, offering reliability in demanding settings.


Advanced Software Integration
The included MadgeTech 4 Software enhances data management with features such as graph views, tabular data presentation, and support for lethality equations. The software facilitates digital calibration, data annotation, and offers full timezone support, making data analysis intuitive and efficient.

Madgetech 4 Graphic Display 

Madgetech 4 is a free downloadable software.  It provides the following software functions:

Data Logger Setup and Configuration
Real Time Display of each Channel
Retrieval of the Recorded Data
Report Generation
Multiple Graph Overlay
Statistics
Zoom In/ Zoom Out
Lethality Equations (F0, PU)
Mean Kinetic Temperature
Data Annotation
Min./Max./Average Lines
Data Logger Calibration


The IFC200 cable is required to connect the data logger to a computer.

.

To Order  click on the model below or scroll to the top of the page then click on ADD TO CART button
PART NUMBERS DESCRIPTION
ThermoVault TC101A Thermocouple-based temperature data logger, thermal insulated enclosure, 24 AWG type K thermocouple (0 °C to +482 °C)
IFC200 USB interface cable

 

 Downloads
ThermoVault Data Sheet
ThermoVault User Guide

 

SPECIFICATIONS

INTERNAL CHANNELS
Temperature Range: -40 °C to +80 °C (-40 °F to +176 °F)
Temperature Resolution: 0.1 °C (0.18 °F)
Calibrated Accuracy: ±0.5 °C (0 °C to 50 °C)

REMOTE CHANNELS
Thermocouple Included: Type K, 36 inch length
Thermocouple Connection: Fixed screw terminal (ST model)
Cold Junction Compensation: Automatic, based on internal channel

MAXIMUM EXPOSURE TIME
(based on initial temperature of 25 °C/77 °F)

Ambient Temperature Maximum Duration in Minutes
100°C (212°F) 136
125°C (257°F) 100
150°C (302°F) 80
175°C (347°F) 66
200°C (392°F) 56
225°C (437°F) 49
250°C (482°F) 44
275°C (527°F) 39
300°C (572°F) 36
325°C (617°F) 32
350°C (662°F) 30



GENERAL
Start Modes: Immediate start, delay start up to 18 months, multiple pushbutton start/stop
Memory: 1,047,552 readings, 523,776 readings in multiple start/stop mode
LEDs: 1 per channel and 2 status LEDs
Reading Rate: 1 reading every second up to 1 reading every 24 hours
Alarm: Programmable high and low limits; alarm is activated when temperature reaches or exceeds set limits
Password Protection: Optional password may be programmed to restrict access to configuration options. Data may be read out without the password.
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Battery Type: 3.6 V lithium battery included, user replaceable
Battery Life: 10 years typical at a 15 minute reading rate
Data Format: Date and time stamped °C, °F, K, °R, μV, mV, V
Time Accuracy: ±1 minute/month at +20 °C (+68 °F), stand alone data logging
Computer Interface: IFC200 USB Interface Cable; 115,200 baud
Operating System Compatibility: Windows XP SP3 or later
Operating Environment: Data Logger: -40 °C to +80 °C (-40 °F to +176 °F), 0% RH to 95 %RH non-condensing
System: -40 °C to +350 °C (-40 °F to +662 °F) time limited, 0 %RH to 95 %RH non-condensing
Dimensions: 5.1 in x 4.5 in x 2.5 in (129 mm x 113 mm x 64 mm)
Weight: 3.5 lbs (1590 g)

Reviews