- Carry 6 Single Function Calibrators In the Palm of Your Hand
- Intuitive EZ-DIAL Double Click Menu Is Easy to Setup
- Use as a Milliamp & Voltage Meter
- Calibrate in C/F: T/C J/K/T/E/R/S/B/N/G/C/D/L/U/P; Pt 100 Ohm (3850, 3902, 3916, 3926) & 1000 Ohm (3850) Copper 10 & 50 Ohm; Nickel 110 & 120 Ohm
- Source & Read Frequency to 2000 CPM, 999.99 Hz, 9999.9 Hz to 20.000 kHz
- Troubleshoot loop & Wiring Problems
- Easily Set Any 3 Outputs Plus Automatic Stepping & Ramping
- Measure Temperature Sensors, Frequency Pickups, Loop Currents & Voltage Levels
- Comes With NIST Certificate, Rubber Boot, Carrying Case, 1 Red, 1 Black Lead w/ Banana to Alligator Clip, 2 Red, 2 Black Leads w/Banana to Spade Lugs, Batteries
Thermocouple Calibrator
The easy to use 820 allows you to calibrate thermocouples directly in C/F. Quickly set any value to within 0.1C with the adjustable digital potentiometer EZ-DIAL plus recall any 3 temperatures for instant recall with the EZ-CHECK switch. Measure thermocouple sensors; trouble shoot sensor connections to find broken wires or corroded connections. Connect your thermocouple with a mini thermocouple connector and the PIE 820 measures the probes in C/F.
RTD, Resistance Calibrator
With the 820 you can check and calibrate all your RTD instruments and measure RTD sensors and show automatic indication of connections on the display for simple hook-ups. The PIE 820 works with the RTDs you use including Platinum 100 (alpha = 3850, 3902, 3916, 3926) & 1000 (alpha = 3850) Ohm, Copper 10 & 50 Ohm, Nickel 100 and 120 Ohm. Easily set any value quickly to within 0.1C with the adjustable digital potentiometer EZ-DIAL plus store any 3 temperatures for instant recall with the EZ-CHECK switch. Or use like a decade box from 0.0 to 401.0 and from 0 to 4010 Ohms. The 820 is compatible with all process instruments. Connect directly to the RTD inputs of smart transmitters, PLCs, DCS and multichannel recorders and verify their outputs or displays. The 820 works with older instruments with fixed excitation currents and newer multichannel instruments that switch the excitation current between input channels. And with the PIE 820 you can measure RTD sensors and trouble shoot sensor connections to find broken wires. Connect 2, 3 or 4 wires RTDs and the 820 measures the RTD in C/F.
Milliamp Calibrator
Check, calibrate and measure all your current signal instruments in a 4 to 20 milliamp DC loop. Access any point in the loop. Source & Read 0.00 to 24.00 mA. Simulate a 2 wire transmitters or use the 820 to simultaneously power your 2 wire transmitter and measure its output. Source milliamps; calibrate recorders, digital indicators, stroke valves or any instrument that get their input from a 4 to 20 mA loop. Easily set any value quickly to within 0.01 mA with the adjustable digital potentiometer EZ_DIAL or use preset 4.00 mA (0.0%) and 20.00 mA (100.0%) EZ-CHECK settings. The 820 can read loop current, calibrate using loop power and power and measure 2 wire transmitters.
Voltage Calibrator
You can check, calibrate and measure all your voltage, millivolt and pH signal instruments. Source 0 to 10.25 V dc and -10.00 to 80.00 mV. Read DC volts; You can measure from 10.25V, -10.00 to 80.00mV and 0.0 to 60.0 VDC. Use the PIE 820 to check loop power supplies, I/V converters, 1 to 5 volt signals.
Frequency Calibrator
Calibrate flow meters and frequency instruments; generate zero crossing square wave to check, calibrate and measure all the frequency signal instruments in your facility. Source and read frequencies from 1 to 2000 CPM, 0.01 to 999.00Hz, 0.1 to 9999.9 Hz and 0.001 to 20.000kHz. The PIE 820 has a green LED that flashes in sync with the output frequency. Select a frequency and hold the calibrator up to the optical sensor.
Continuity Checker
Troubleshoot wiring and connection problems; use the built-in continuity checker to look at wiring and connections during installation or to locate shorts. 820 will beep from 0.100 Ohms.
To Order the model below scroll to the top of the page then click on ADD TO CART button |
PART NUMBER | DESCRIPTION |
CL-PI-820 | PIE 6 Function Process Calibrator |
Datasheets and Operating Instructions |
SPECIFICATIONS
READ mA
Ranges and resolution, 0.00 to 24.00 mA or -25.0 to 125% of 4-20 mA
Accuracy, < +/- (0.03% of full scale)
Voltage burden, < 2V at 24 mA
Overlook/current limit protection, 25 mA nominal
SOURCE mA/POWER & MEASUREMENT 2 WIRE TRANSMITTERS
Ranges and resolution, 0.00 to 24.00 mA or -25.0 to 125.0% of 4-20 mA
Accuracy, < +/- (0/03% of full scale)
Loop compliance burden, > 24 DCV @ 20.00mA
Loop drive capability, 1200 Ohms at 20 mA for 15 hrs nominal
mA 2-WIRE TRANSMITTER SIMULATION
Accuracy, Same as source/power & measure
Voltage burden, < 2V at 20 mA
Overload/current limit protection, 24 mA nominal
Loop voltage limits, 2 to 60 VDC (fuse-less protected fro reverse polarity connections)
DC VOLTAGE READ
Range and accuracy, 0.00 to 80.00 mV, < +/- (0.03% of full scale)
Input resistance, > 1MOhm
SOURCE V dc
Range and accuracy, -10.00 to 80.00 mV, < +/- (0.03% of full scale), 0.00 to 10.25V, < +/- (0.03% of full scale +0.005V)
Source current, >24 mA
Sink current, >16 mA
Output impedance, <1 Ohm
Short current duration, infinite
THERMOCOUPLE SOURCE
Accuracy, +/-(0.03% of full scale) (full scale being 80.00mV)
Cold junction compensation, +/-0.1%C (thermistor traceable to NIST for 11 years)
Output impedance, <1 Ohm
Source current, >20 mA (drives 80 mV into 10 Ohms)
THERMOCOUPLE READ
Accuracy & cold junction compensation, same a t/c source
Input impedance, > 1 megohms
Open t/c threshold: pulse, 10K Ohms < uamp pulse for 300 milliseconds (nominal)
RTD, OHMS & CONTINUITY READ
Resistance ranges, 0.0 to 401.0 to 4010 Ohms
Accuracy, +/-(0.03% of full scale + 0.075 Ohms)
Excitation current, 1.0 mA to 401 Ohms, 0.5 mA to 4010 Ohms (nominal)
Continuity, 0.0 to 401.0 Ohms; beeps from 0.0 to 100 Ohms
RTD & OHMS SOURCE
Accuracy from 1 to 10.2 mA external excitation current, +/-(0.03% of full scale + 0.075 Ohms)
Accuracy from below 1 mA of external excitation current, +/-(0.03% of full scale +0.075 Ohms +0.025 mV/mA excitation current
Resistance ranges, 0.0 to 401.0 to 4010 Ohms
Allowable excitation current range, <401 Ohms: 10.2 mA max; steady or pulsed/intermittent, 402 to 4000 Ohms: 1 mA max; steady or pulsed/intermittent
Pulsed excitation current compatibility, DC to 0.01 second pulse width
FREQUENCY SOURCE
Ranges, 1 to 2000 CPM, 0.01 to 999.99 Hz, 0.1 to 9999.9 Hz
Accuracy, +/-(0.03% of full scale)
Output waveform, square wave, zero crossing -1.0 to +5V peak-to-peak +/-10%
Risetime (10 to 90% of amplitude, < 10 microseconds
Output impedance, <1 Ohm
Source current, >1 mA rms at 20 Hz
Sort circuit duration, infinite
Optical coupling, green LED (HZ SYNC) flashes at output frequency
FREQUENCY READ
Ranges and accuracy, sames a frequency source
Accuracy, +/-(0.03% of full scale)
Trigger level, 1 V rms, dc coupled
Input impedance, > 1 Meg Ohm + 60 pF
OPERATING TEMPERATURE RANGE: -20 to 60C (-5 to 140F)
STORAGE TEMPERATURE RANGE: -30 to 60C (-22 to 140F)
TEMPERATURE EFFECT: < +/- 0.01%/C of full scale
RH RANGE: 10% <RH <90% (0 to 35C) non-condensing, 10% <RH< 70% (35 to 60C) non-condensing
NORMAL MODE REJECTION: 50/60 Hz, 50 dB
COMMON MODE REJECTION: 50/60 Hz, 120 dB
NOISE: < /- less significant digit from 0.1 to 10 Hz
DIMENSIONS: 5.63 x 3.00 x 1.60 in (143 x 76 x 41mm)
WEIGHT: 12.1 oz (0.34 kg)
BATTERIES: 4-AA (supplied)
BATTERY LIFE: read function > 20 hrs; source mA > 14 hrs @ 12 mA into 250 Ohms; pwr/meas mA > 12 hrs at 20 mA, source V, Ohm, T/C, RTD & Hz > 20 hrs
LOW BATTERY: 1 hr of operation left
PROTECTION AGAINST MISCONNECTION: over-voltage protection to 60 vrms (rated for 30 seconds), red LED indicates OVERLOAD or out of range conditions
DISPLAY: high contrast graphic LCD with 0.315" (8.00 mm) LED backlight for use in low lit areas