- Easy To Use, Dependable Milliamp Loop Calibrator
- Source Milliamps
- Calibrate Using Loop Power
- Read Loop Current
- Power and Measure 2 Wire Transmitters
- Read DC Volts
- Comes With Rubber Boot, NIST Certificate, Test Lead w/Alligator Clips, Batteries
With the 334 you can check, calibrate and measure all your current signal instruments in a 4 to 20 milliamp DC loop. The PIE 334 can be used at any access point in your loop. Source and read 0.00 to 24.00 mA, simulate a 2 wire transmitter or use the meter to simultaneously power your 2 wire transmitter and measure its output. Calibrate in confidence: accurate to +/-0.025% of span at 4.00 & 20.00 EZ-CHECK outputs.
Source milliamps: calibrate recorders, digital indicators, stroke valves or any instruments that get their input from a 4 to 20 mA loop. Easily set any value quickly to within 0.01 mA with the adjustable digital potentiometer DIAL or use the preset 4.00 mA (0.0%) and 20.00 mA (100%) EZ-CHECK settings.
With the 334 you can calibrate using loop power. Check loop wiring and receivers by using the 334 in place of a 2 wire transmitter. PIE 334 uses any loop power from 2 to 100 V DC.
Read loop power with the 334. Check controller outputs or measure the milliamp signal anywhere in the loop. The 334 measures 0.00 to 52.00 mA (-25.0 to 300.0%) signals with greater accuracy than a typical multimeter.
Power and measure 2 wire transmitters with the 334. You can simultaneously output 24V DC to power any and all devices in a process loop using the internal batteries and internal switching power supply while measuring the output of a 2 wire transmitter with any other loop meters.
Read DC volts with the 334. You can measure from -99.99 to +99.99 VDC with 0.01 volt resolution. Use 334 to check loop power supplies, I/V converters, 1 to 5 volt signals and other voltages making it unnecessary to carry a multimeter.
To Order the model below scroll to the top of the page then click on ADD TO CART button |
PART NUMBER | DESCRIPTION |
CL-PI-334 | PIE 334, 4-20 mA Loop Calibrator |
Datasheets and User Manuals
PIECAL 334 Operating Instructions
SPECIFICATIONS
READ mA
Ranges and resolution, 0.00 to 52.00 mA or -25.0 to 300% of 4-20 mA
Accuracy below 24.010 mA, < +/-(0.05% of 24.00 mA) (+/-0.01 mA), accuracy above 24.010mA, < +/-(0,05% of 52.00 mA) (+/-0.3 mA)
Voltage burden, < 2V at 50 mA
Overload/current limit, 54 mA nominal
Battery life, >125 hrs nominal
SOURCE/POWER & MEASURE 2 WIRE TRANSMITTERS
Ranges and resolution, 0.00 to 24.00 or -25.0 to 125.0% of 4-20 mA
Accuracy EZ-CHECKS at 4 & 20 mA, < +/- (0.025% of span at 4 mA & 20 mA) (+/-0.005mA), EZ-CHECKS at 0.000 to 24.000 mA, < +/- (0.05% of 24 mA span) (+/-0.012mA)
Noise, < +/- 1/2 least significant digit
Temperature effect, < +/- 0.005% C of FS
Loop compliance voltage, > 24 DCV @ 20.00 mA
Loop drive capability, 1200 Ohms at 20 mA for 15 hrs nominal
Battery life, > 30 hrs at 12 mA nominal
2-WIRE TRANSMITTER SIMULATION
Accuracy, sames as source/power & measure
Voltage burden, < 2V at 20 mA
Overload/current limit protection, 24 mA nominal
Loop voltage limits, 2 to 100 VDC (fuse-less protected from reverse polarity connections)
Battery life, > 125 hrs nominal
VOLTAGE READ
Range and resolutions, -99.99 to +99.99 VDC full span (FS)
Accuracy, < +/-0.05% of FS
Temperature effect, < +/- 100 ppm/C of FS
Input resistance, > 2 MOhms
Battery Life, > 125 hrs nominal
OPERATING TEMPERATURE RANGE: -20 to 60C (-5 to 140F)
STORAGE TEMPERATURE RANGE: -30 to 60C (-22 to 140F)
RH RANGE: 10% <RH <90% (0 to 35C) non-condensing,10% <RH <70% (35 to 60C) non-condensing
DIMENSIONS: 5.63 x 3.00 x 1.60 in (143 x 76 x 41mm)
WEIGHT: 12.1 oz (0.34 kg) including boot and 4-AA batteries
LOW BATTERY: Low battery indication with nominal 1 hr of operation left
PROTECTION AGAINST MISCONNECTION: Over-voltage protection to 135 vrms (rated for 30 seconds) or 240 vrms (rated for 15 seconds
DISPLAY: High contrast LCD with 0.315" (8.0 mm) high digits