What is the total error associated for the thermocouple measurement using TCTempX data logger?
TCTempX - Multi Channel Thermocouple Loggers, 4, 8, 12 and 16 Channel
Status:
Open Oct 06, 2024 - 05:47 PM
Sales
1answer
Answer
Oct 06, 2024 - 05:47 PM
The accuracy stated for TCTempX is for data logger alone and does not include the thermocouple accuracy. This is typical for most instruments since thermocouple accuracy varies based on the thermocouple class. In the United States there are two common thermocouple accuracy classes. They are the standard limits error and special limits of error accuracy classes. The total system accuracy includes both the TCTexmpX accuracy and the thermocouple accuracy.
View More Product Details
TCTempX - Multi Channel Thermocouple Loggers, 4, 8, 12 and 16 Channel
Up to 4 Hz Reading Rate Memory Capacity of 2,000,000+ Open Channel Detection Accepts Thermocouple Types - J, K, T, E, R, S, B, N External Power or User Replaceable Battery Pushbutton or Programmable Start ISO17025 Calibration Certificate included Free Data Collection, Display and Analysis Software T...